Gas analysis in the pressure range of < 10-4 mbar
QMG 422 and QMG 700 mass spectrometers are designed for qualitative and quantitative analysis. They are characterized by the highest measurement stability and sensitivity. Mass spec. system with Analyzer QMA 410: 16 mm molybdenum rod system, highest resolution (< 1 ppb), mass ranges up to 340 amu. With Analyzer QMA 400: 8 mm molybdenum rod system, high resolution (10 ppb), mass ranges up to 2048 amu. With Analyzer QMA 430: 8 mm stainless steel rod system, high resolution, mass range 1-300 amu.
- QMG 422 with QMA 410 with 16 mm molybdenum rod system
- QMG 422 with QMA 430 with 8 mm stainless steel rod system
- QMG 422 with QMA 400 with 8 mm molybdenum rod system
- QMG 700, HiQuadTM with QMA 410, 16 mm molybdenum rod system
- HiQuadTM, QMG 700, with QMA 430, 8 mm stainless steel rod system
Applications
- All mass ranges between 128 and 2048 amu
- Qualitative gas analysis up to 10-4 mbar
- For detection of neutrals, positive and negative ions
Customer value
- Modular, adaptable, expandable
- Higher resolution selectable, even in connection with low mass numbers, e.g. for separating He and D2
- Satisfies analytical requirements: Detection in the ppb range, high reproducibility and long-term stability, minimal contribution to neighboring masses, high dynamics